• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING DEXCOM SHARE SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING DEXCOM SHARE SYSTEM; MDS Back to Search Results
Model Number MT22608
Device Problem Improper Device Output (2953)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient's mother contacted dexcom on (b)(6) 2015 to claim no audio output on (b)(6) 2015.There was no report any injury or medical intervention.No additional event or patient information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defects were found.A "try it" manual test was performed and speaker did not sound.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING DEXCOM SHARE SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5321446
MDR Text Key34916294
Report Number3004753838-2015-74168
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000187
UDI-Public(01)00386270000187(241)MT22608(10)5198373(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22608
Device Catalogue NumberSTK-PR-001
Device Lot Number5198373
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 MO
Patient Weight16
-
-