Model Number SC-3138-35 |
Device Problem
Cut In Material (2454)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 12/02/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Additional suspect medical device component involved in the event: model #: sc-3138-35, serial #: (b)(4), description: scs phiii ext 35cm.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation.The patient underwent a revision procedure wherein the ipg was replaced.No device malfunction was suspected with the ipg.During the procedure while the extensions were removed from the lead, it was noted that the lead looked weird as it had a very wide space.The very end of the lead contact was cut off, and appeared to be damaged.The physician could not verify if the issue was from the device.However, when the physician plugged the lead into the header, everything lined up well.The patient was doing well postoperatively.No further course of action.
|
|
Manufacturer Narrative
|
(b)(4).Additional suspect medical device component involved in the event: model #: sc-3138-35, serial #: (b)(4), description: scs phiii ext 35cm.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation.The patient underwent a revision procedure wherein the ipg was replaced.No device malfunction was suspected with the ipg.During the procedure while the extensions were removed from the lead, it was noted that the lead looked weird as it had a very wide space.The very end of the lead contact was cut off, and appeared to be damaged.The physician could not verify if the issue was from the device.However, when the physician plugged the lead into the header, everything lined up well.The patient was doing well postoperatively.No further course of action.
|
|
Search Alerts/Recalls
|