• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY TULSA DENTAL SPECIALTIES LEXICON K-FILE; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY TULSA DENTAL SPECIALTIES LEXICON K-FILE; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number KFS00825
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 11/23/2015
Event Type  malfunction  
Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude injury or illness that would necessitate medical or surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that two k-files broke during treatment on the same patient, both in the same tooth.The separated pieces were left in the canal.The doctor filled the canal with gutta percha leaving the file in the tooth as part of the fill.
 
Manufacturer Narrative
No used file was returned.9 unused files were sent in for investigation.Unused product has been evaluated and was found in compliance with specifications.Nothing unusual to report was found during dhr review.Root causes are not identified.This kind of event is tracked and we monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEXICON K-FILE
Type of Device
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
DENTSPLY TULSA DENTAL SPECIALTIES
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM   81737
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5323600
MDR Text Key35014230
Report Number9611053-2015-00056
Device Sequence Number1
Product Code EKS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberKFS00825
Device Lot Number130905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-