• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES BI-PHASIC
Device Problems Pacing Problem (1439); Inappropriate or Unexpected Reset (2959)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
 
Event Description
Complainant alleged that during biomed testing the device's pacer output resets to zero.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a loose hex nut on the control board.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5324982
MDR Text Key34916945
Report Number1220908-2015-03331
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Device Lot NumberN/A
Other Device ID Number00847946002916
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-