Model Number MSERIES BI-PHASIC |
Device Problems
Pacing Problem (1439); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during biomed testing the device's pacer output resets to zero.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a loose hex nut on the control board.Analysis for reports of this type has not identified an increase in trend.
|
|