MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 7078396
Device Problem Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).This part is not approved for use in the united states; however a like device catalog # 5540030, 510k # (b)(4) was cleared in the united states.Neither the device nor applicable imaging films were returned to the manufacturer for evaluation, therefore, the cause of the event cannot be determined.
 
Event Description
It was reported that patient underwent posterior fusion at th10-s2ai.During surgery, burrs came off while inserting the set screw into the mas screw due to cross-threading and set screw was replaced.Patient complications are reported unknown.
 
Manufacturer Narrative
Product analysis:visual and microscopic examination of the returned set screw identified the upper flanks of the set screw surface is skived, starting at the start of the thread, with a thin portion of the upper flank removed.This damage is consistent with overloading the set screw thread due to incomplete seating of the rod in the mas saddle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5331379
MDR Text Key35037515
Report Number1030489-2015-03583
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/20/2023
Device Catalogue Number7078396
Device Lot NumberH5164179
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-