MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
Catalog Number 7078396 |
Device Problem
Metal Shedding Debris (1804)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/15/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).(b)(4).This part is not approved for use in the united states; however a like device catalog # 5540030, 510k # (b)(4) was cleared in the united states.Neither the device nor applicable imaging films were returned to the manufacturer for evaluation, therefore, the cause of the event cannot be determined.
|
|
Event Description
|
It was reported that patient underwent posterior fusion at th10-s2ai.During surgery, burrs came off while inserting the set screw into the mas screw due to cross-threading and set screw was replaced.Patient complications are reported unknown.
|
|
Manufacturer Narrative
|
Product analysis:visual and microscopic examination of the returned set screw identified the upper flanks of the set screw surface is skived, starting at the start of the thread, with a thin portion of the upper flank removed.This damage is consistent with overloading the set screw thread due to incomplete seating of the rod in the mas saddle.
|
|