• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Event Description
It was reported that during a cryo ablation procedure fluid leaked out of the valve of the sheath.The sheath was replaced and the procedure was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the flexcath advance sheath 4fc12 with lot number 75953-97 was returned and analyzed.Visual inspection showed the device was intact with no apparent issues.Air aspiration was reproduced with introduction of a test balloon catheter.Dissection of the device showed the hemostatic valve was leaking.In conclusion, the reported issue of a leaking hemostatic valve has been confirmed through testing.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5335814
MDR Text Key35217470
Report Number3002648230-2015-00478
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/15/2017
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number75953
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-