Model Number 4FC12 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/10/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Event Description
|
It was reported that during a cryo ablation procedure fluid leaked out of the valve of the sheath.The sheath was replaced and the procedure was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the flexcath advance sheath 4fc12 with lot number 75953-97 was returned and analyzed.Visual inspection showed the device was intact with no apparent issues.Air aspiration was reproduced with introduction of a test balloon catheter.Dissection of the device showed the hemostatic valve was leaking.In conclusion, the reported issue of a leaking hemostatic valve has been confirmed through testing.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|