Catalog Number 6605526
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2015
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument on 12/07/2015.The fse found that the manual aspiration probe was bent and was causing the errors.The fse replaced the probe, which repaired the errors.The fse also found that the autoloader cassette was not properly advancing.The fse replaced the cassette assembly to repair the issue.The repairs were verified by the fse.(b)(4).
 
Event Description
The customer reported the coulter hmx analyzer with autoloader was generating backwash errors.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Manufacturer Narrative
The device date of manufacture was changed from 09/01/2006 to 06/01/2009.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER HMX ANALYZER WITH AUTOLOADER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key5337389
MDR Text Key34755016
Report Number1061932-2015-01898
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K010765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605526
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-