Model Number 3100A |
Device Problems
Failure to Power Up (1476); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/03/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Carefusion complaint number (b)(4).In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.At this time, carefusion has not received the device from the customer.(b)(4).
|
|
Event Description
|
The customer stated that the start/stop button is not engaging well when pressed and the centering display is only showing 2 leds lit.This occurred during set-up therefore there was no patient involvement.
|
|
Manufacturer Narrative
|
The carefusion field service representative evaluated the unit on (b)(6) 2015 and found the issue to be that the pressure regulator was out of specification.The driver was tested a various settings and no further issues were found.The ventilation was run through the alarms test, the patient circuit calibration test and performance checks and all passed.The ventilator was placed back into service.
|
|
Search Alerts/Recalls
|