• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4471R
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, one increased intra-peritoneal volume (iipv) event was identified.This event occurred in the therapy initiated on (b)(6) 2015 at 21:31:31.During night drain cycle five, the patient's ultrafiltration reading was 1392ml, indicating the home patient drained 1392ml more than their maximum programmed fill volume of 2300ml.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received and the evaluation is complete.During the event history log review, an increased intra-peritoneal volume (iipv) event was identified.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the event.The homechoice device received a returned instrument testing evaluation (rite).This evaluation includes functional and electrical testing on the device.Upon conclusion of the investigation, the cause of the iipv event was one or more cycles advanced to the next fill when a slow or no flow occurred above the minimum drain volume threshold.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5339270
MDR Text Key34883328
Report Number1416980-2015-46057
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5C4471R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age61 YR
Patient Weight81
-
-