• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Overfill (2404)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, one increased intra-peritoneal volume (iipv) event was identified.This event occurred in the therapy initiated on (b)(6) 2015 at 00:04:25.During night drain cycle four, the patient's ultrafiltration reading was 2346ml, indicating the home patient drained 2346ml more than their maximum programmed fill volume of 2500ml.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received and the evaluation is complete.During the event history log review, an increased intra-peritoneal volume (iipv) event was identified.The homechoice device received a returned instrument testing evaluation (rite). this evaluation included functional and electrical testing of the device.A visual inspection was performed.A service history review revealed no indication that the parts replaced during servicing caused or contributed to this event.Upon conclusion of the investigation, the cause of this issue was determined to be one or more cycles advances to the next fill when slow or no flow occurred above the minimum drain volume threshold.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5339380
MDR Text Key35247820
Report Number1416980-2015-46093
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age25 YR
Patient Weight91
-
-