BAXTER HEALTHCARE - MOUNTAIN HOME UV FLASH TRANSFER SETS; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number T5C4326 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/04/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received.Should additional information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported it had been difficult to set the transfer set into uv assist device.There was gap between actual sample of transfer set and cap.After changing to new transfer set, there were no issues.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The device was received for evaluation.A gross visual inspection and microscopic inspection were performed and identified a damaged patient adaptor and a gap between the transfer set's flow control barrel and the disconnect cap.Functional testing of the device included leak testing, clear passage testing, and clamp function testing; no issues were found that could have caused or contributed to the reported problem.The reported issue was verified; however, the cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|