• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; BREATHING SYSTEM HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; BREATHING SYSTEM HEATER Back to Search Results
Catalog Number 5705
Device Problems Insufficient Heating (1287); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device sample was not returned for evaluation the time of this report.
 
Event Description
The customer alleges that the unit is no longer heating.No patient injury/harm reported.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the heater was received without accessories.The unit is approx.618 days of age and out of warranty.Upon evaluation of unit, it was found that the heating element was damaged, and no longer produced heat.Based on the investigation performed, the reported complaint was confirmed.The heater will be repaired and returned to the customer.
 
Event Description
The customer alleges that the unit is no longer heating.No patient injury/harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON THERMAGARD NEBULIZER HEATER
Type of Device
BREATHING SYSTEM HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
2917 weck drive
research triangle park NC 27709
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5340053
MDR Text Key34918622
Report Number1044475-2016-00001
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5705
Device Lot Number18989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/28/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-