It was reported a 25mm gore cardioform septal occluder was selected to close an atrial septal defect static measured 8mm and balloon sized to 11mm.The device was successfully implanted at which time a 15mm gore cardioform septal occluder was selected to close a second defect static measured to 5mm.Several attempts were made but the device could not be advanced through the defect.During attempts to implant the second device, the first device embolized to beneath the mitral valve.The patient was taken to surgery, the device was successfully removed and the defects repaired.The patient was doing well following surgical closure of the defects.
|