The device was not returned to the nidek.However, the device was evaluated by the nidek field service engineer fse at the facility on 12/15/2015.Fse tested and evaluated the device.Complete operational check was performed and device passed the tests performed.Fse also checked the laser beam profile using the polarized film at 24kv, 30kv and 1hz and found weak center within profile.Fse advised to replace the front and the rare optics to correct reported problem.After interview with the technician fse found out that the customer is using third party to perform the preventive maintenance service on the device.Last preventive maintenance was performed in early (b)(6) 2015 by the third party.Fse recommended customer that the laser should be maintained by the qualified nidek inc.Service personnel.(reference: as per the operator's manual for ec-5000; 3: safety; 3.7: maintenance and check: only nidek service personnel should disassemble and repair the system.) nidek clinical specialist contacted technician to gather additional patient information on 12/22/2015.In total three patients had undercorrection and all the patients were in healing phase.Nidek inc.Considers it a reportable event as the device has malfunctioned and has a potential to cause or contribute to a serious injury if the malfunction were to recur.Separate mdr will be submitted for all three patients.Please refer the table below for additional details.Age: 44yrs, sex: male, date of surgery performed: (b)(6) 2015, mdr #: 3002807715-2016-00001.Age: 47 yrs, sex: male, date of surgery performed: (b)(6) 2015, mdr #: 3002807715-2016-00002.Age: 20 yrs, sex: female, date of surgery performed: (b)(6) 2015, mdr #: 3002807715-2016-00003.
|