The infusion pump was not manufactured by ekos.There are possible reasons for the iv pump alarm.These include: the pressure setting (pump psi) is set too low, aspiration of blood into the drug line, kinking the line due to patient positioning or patient movement, or suturing down the line such that it becomes occluded.A nurse at the site told an ekos sales representative that the drug catheter had been tied down with a suture by a hospital technician.The device was not returned for evaluation, so ekos cannot determine the cause of the pump alarm.The ekosonic mach4 endovascular device instructions for use, clearly state to prepare the catheters outside of the body by flushing them prior to placement in the patient.In addition, the instructions: warn against drawing blood back into the drug lumens, as blood may occlude the drug holes.Despite good faith attempts, ekos was unable to obtain additional information.Device was not returned for evaluation.
|