BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number N/A
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Not Applicable (3189)
Event Date 07/17/2015
Event Type  malfunction  
Manufacturer Narrative
The fse (field service engineer) reported that the vcs (volume conductivity scatter) analog 5 card failed prematurely causing high rf (radio frequency) and that no results were being generated.The fse replaced the analog 5 card to resolve the issue.The repairs were verified per established procedure.(b)(4).
 
Event Description
The fse (field service engineer) reported premature part failure on the coulter lh750 hematology analyzer.The radio frequency was higher and no results were being generated when the problem was noticed.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami 33196
3053802031
MDR Report Key5357279
MDR Text Key35551380
Report Number1061932-2015-01852
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number6605632
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-