BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY
|
Back to Search Results |
|
Model Number H802228240022 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/16/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Age at the time of event: 18 years or older.(b)(4).Device was returned for analysis.Visual inspection was performed and noted that the body was kinked at 19.1cm and at 171.5cm approximately from the proximal end; moreover, the distal end is broken at 171.5cm approximately from the proximal end.Evidence of wear reduction was noted on the broken section.The most distal section including the spring tip was not returned.All the outer diameter measurements are within the specifications.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on 14dec2015.It was reported that a guidewire was kinked.A 330cm rotawire was selected to advance to the target lesion.During preparation, it was noted that the guidewire was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, device analysis revealed a broken distal end of the guidewire.
|
|
Search Alerts/Recalls
|
|
|