• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, BIO-CORKSCREW FT; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, BIO-CORKSCREW FT; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number AR-1927BCF-1
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 11/16/2015
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested/is expected for evaluation but has not yet been received.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Based on the information provided, the most likely cause(s) of this event include improper bone preparation for the hardness of bone encountered, not inserting the implant co-axial to the bone tunnel, prying/leveraging the driver while the implant is still loaded, and/or over-insertion.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.Device expected but not yet returned.
 
Event Description
It was reported that the anchor is broken at its distal portion.An attempt to remove the fragments were made but fragments may still be in the patient.A second anchor had to be used to finish the case.No further infiormation was made available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, BIO-CORKSCREW FT
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key5360463
MDR Text Key35670959
Report Number1220246-2015-00391
Device Sequence Number1
Product Code MAI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K082810
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/01/2017
Device Catalogue NumberAR-1927BCF-1
Device Lot Number1491586
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-