A medtronic representative reported that, while in a laser induced thermal therapy procedure, after the first tmap was over, they were unable to close the session.In trouble-shooting, they were unable to re-start computer- server was busy.A hard re-boot of the laser induced thermal therapy system restored normal function.As a result of hard re-boot the first session, laser powers were not available as logs.After re-boot system functioned normally.The surgeon completed the procedure with the use of the laser induced thermal therapy system.Delay in therapy was less than one hour.There was no impact on patient outcome.
|