During a standard follow-up on (b)(6) 2016, it was verified 3 warning messages: "[3] low shock impedance.Defibrillation system ineffective." "[48] max shock energy ineffective on (b)(6) 2016" "[46] high shock impedance detected on (b)(6) 2016: defibrillation system ineffective." nine shocks were delivered, all of them with stored energy = 41.9j and delivered energy = 0j.The shock lead integrity trend is stable between the expected values.In the memories of the device it was also notified atrial undersensing.The physician changed atrial sensibility from 0.4 mv to 0.2mv.It was scheduled a new follow-up to (b)(6) 2016.
|