SYNTHES OBERDORF OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.021 |
Device Problems
Poor Quality Image (1408); Device Inoperable (1663); Defective Component (2292)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that device did not run at all, the housing was discolored and the bearings were defective.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to be normal wear from use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the oscillating saw attachment device did not run at all and had defective bearings and discolored housing.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|