• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS NAMIC; STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS NAMIC; STOPCOCK Back to Search Results
Catalog Number H965700350091
Device Problem Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Date 12/21/2015
Event Type  malfunction  
Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The angiodynamics (b)(6) 2015 complaint report was reviewed for the product family of stopcocks/manifolds and the failure mode, "hole in pouch." no adverse trends were identified.The return sample was visually inspected and the hole in the pouch was confirmed.As part of the receiving process at (b)(4) (the distributor in (b)(4)), all pouched products are removed from their inner boxes and a (b)(4) label (in (b)(4)) is applied to each pouch.Additional handling may occur if the pouches are 100% visually inspected.The pouched product is then re-boxed into the inner box by the (b)(4) warehouse employees.The hole in the tyvek was likely caused by the handle of the stopcock, however, what caused the handle to be pushed against the tyvek in a manner that resulted in a hole cannot be determined.Potential contributing factors included: handling in the in-process bins afer sealing but prior to being sent to the final box area or handling of the pouches as they are placed in the inner boxes at angiodynamics.Handling during transit to the warehouse.Handling at (b)(4) during the labeling/inspection and re-boxing process.All pouches are 100% inspected per angiodynamics procedures during the sealing and final box processes.(b)(4).
 
Event Description
As reported by angiodynamic's distributor in (b)(6), in the distributor's warehouse a small hole was found in the tyvek portion of the stopcock pouch, breaching the sterility.The device had not been provided to a hospital and was returned to angiodynamics for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAMIC
Type of Device
STOPCOCK
Manufacturer (Section D)
ANGIODYNAMICS
glens falls NY
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
law ryan
10 glens falls technical park
glens falls, NY 12801
5187424488
MDR Report Key5366327
MDR Text Key36090808
Report Number1317056-2016-00006
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K782095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 12/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2018
Device Catalogue NumberH965700350091
Device Lot Number4946606
Other Device ID NumberSTOPCOCK
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/29/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-