Model Number F110
Device Problems Device Alarm System (1012); Premature Discharge of Battery (1057)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/20/2015
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this device was emitting tones.Upon interrogation it was noted that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.No adverse patient effects were reported.The device was explanted and is not expected to be returned for analysis.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to a product performance anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was included in the advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELIGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5371218
MDR Text Key36207369
Report Number2124215-2015-14836
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/21/2010
Device Model NumberF110
Other Device ID NumberTELIGEN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/23/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0088-2015
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age72 YR
-
-