• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number NOMAD PRO
Device Problem Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2015
Event Type  malfunction  
Manufacturer Narrative
The battery handset and charging cradle were returned and undergoing investigation.A follow-up report will be submitted upon the investigation completion.
 
Event Description
A local aribex distributor informed us that a nomad pro handset had partially melted on the charging station.There were no injuries reported.
 
Manufacturer Narrative
The analysis of the returned device has been completed.Our records indicate that the device was assembled in july 2012.The handset was manufactured in october 2012.The returned unit was inspected visually and by x-ray.The investigation found evidence of heavily applied dental office cleaners and fluid ingress into the interior of the handset.It was determined that there was a short circuit in the upper and lower battery pack circuit boards that lead to the battery failure.No injuries were reported.The nomad pro operator manual provides instructions on the proper care, cleaning, and maintenance of the device.This concludes our investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key5371399
MDR Text Key36462300
Report Number1017522-2016-00002
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Dental Assistant
Device Model NumberNOMAD PRO
Device Catalogue Number0.850.00009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-