Catalog Number 2101-0200
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 12/23/2015
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that during a right total hip procedure, the shell inserter handle broke.This happened after impaction and case completed without any delay or adverse consequence to patient or user.
 
Manufacturer Narrative
An event regarding fractured handle involving a universal impactor/positioner was reported.The event was confirmed.Method & results: device evaluation and results: the returned device was in used condition.The visual inspection confirmed the fractured handle.Medical records received and evaluation: no medical records were received.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other similar events reported for this lot.Conclusions: the investigation concluded the handle fractured due to a design issue.
 
Event Description
It was reported that during a right total hip procedure, the shell inserter handle broke.This happened after impaction and case completed without any delay or adverse consequence to patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5378336
MDR Text Key36674908
Report Number0002249697-2016-00151
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/30/2015
Device Catalogue Number2101-0200
Device Lot NumberB5AJR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-