• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125250-12
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/30/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat an unspecified coronary artery.During loading of the 2.5 x 12 mm xience alpine stent delivery system (sds) onto the guide wire, at about 30-50 cm in, the guide wire exited the side of the sds shaft before reaching the guide wire exit notch.The sds did not go into the anatomy.The device was removed from the guide wire and a new stent delivery system was used to complete the case.No adverse patient effects or clinically significant delay in the procedure were reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: (b)(4).The device was returned for analysis.The tear in the balloon was able to be confirmed.Based on a visual and functional inspection of the returned device, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5381698
MDR Text Key36626495
Report Number2024168-2016-00382
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/10/2018
Device Catalogue Number1125250-12
Device Lot Number5060161
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/12/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-