The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply is not charging the freedom onboard batteries in the freedom driver that was supporting a patient.The customer also reported that the patient was provided with a replacement ac power supply.There was no reported adverse patient impact.The freedom hospital ac power supply was returned to syncardia for evaluation.The hospital ac power supply met the functional and electrical test acceptance criteria.There was no evidence of a blinking led on the hospital ac power supply when it was connected to a power adaptor.The hospital ac power supply was connected to a properly functioning test battery charger with four freedom onboard batteries that each had only one bar illuminated on the charge level indicator.The led on the ac power supply remained illuminated and did not exhibit any blinking during the charging process.All four onboard batteries charged to full capacity.The customer-reported issues could not be duplicated.The hospital ac power supply performed as intended.Despite the customer-reported issues, risk to the patient was low because the freedom driver would not have been prevented from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a backup ac power supply.Because the freedom hospital ac power supply did not meet the visual inspection acceptance criteria, it was taken out of service.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|