SMITHS MEDICAL UNIPERC ADJUSTABLE FLANGE TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/897/080
Device Problems Detachment Of Device Component (1104); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2015
Event Type  malfunction  
Manufacturer Narrative
The device is currently being evaluated; the manufacturer will file a follow-up report detailing the results of the evaluation once it is completed.
 
Event Description
It was reported that during use the moulded joint between the universal adaptor and the inner fixed white tube slipped out.Tracheostomy tube replacement was required.No adverse health outcome resulted from this event.
 
Manufacturer Narrative
One used sample was received for investigation.The device examination confirmed that the connector was disconnected from the tube.The examination also showed that a portion of the molded tube was missing (inflation line holder).The physical characteristics of the returned device (connector broken off and inflation line holder) were determined consistent with excessive force having been applied to the connector during use.The device instructions for use include the warning: "during and after attachment of the breathing system to the tracheostomy tube connector, avoid application of excessive rotational or linear forces on the tube to prevent accidental disconnection or occlusion.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIPERC ADJUSTABLE FLANGE TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL
boundary road
hythe, kent CT216 JL
UK  CT216JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, CT21 6JL
UK   CT21 6JL
Manufacturer Contact
michele seliga
1265 grey fox rd.
st. paul, MN 55112
7633833052
MDR Report Key5388884
MDR Text Key36773809
Report Number2183502-2016-00093
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K083031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/897/080
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-