This device referenced in this report has been returned to keymed ltd.For evaluation.The evaluation found that the distal end cap and the distal end cap ring ("ring") were detached from the distal end, and the reported phenomenon thus was confirmed.Evidence of a chemical attack was found on the surface of the insertion tube.The adhesive glue and binding used to secure the ring and the distal end cap to the distal end of the instrument were found to be missing.There were some scratches on the surface of the distal end cap.The two locating stopper pins were intact with no evidence of damage.The manufacturing record of the subject device was reviewed with no abnormality possibly associated with the reported phenomenon.The instrument was last received by olympus service on (b)(4) 2014 and a minor repair to replace the distal cap and the bending rubber.Based on the evaluation, the exact cause of the parts drop-off could not be conclusively determined at this time, but a possible mechanism of the phenomenon was supposed that the adhesive to secure the ring and the distal end cap was deteriorated by chemical attack such as detergent solution and then external force was applied to the distal end cap.Furthermore, position gap between the distal cap and the ultrasonic unit may cause ultrasonic image irregularity.
|