Catalog Number 8065752200 |
Device Problem
Aspiration Issue (2883)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Additional information has been requested and received.(b)(4).
|
|
Event Description
|
A doctor of ophthalmology reported no aspiration during a cataract surgery with intraocular lens (iol) implant.At first, the irrigation/aspiration (i/a) tip was changed without success.Then, the cassette was changed and the event resolved.The affected eye was the right eye (od).Per the surgeon the cassette contributed to the event.Additional information has been requested and received.
|
|
Manufacturer Narrative
|
Evaluation summary: review of the device history records indicated the order was built to specification.The customer did not retain a sample for this complaint report; therefore, visual inspection or functional testing could not be conducted.The root cause of the customer's complaint could not be established as a sample has not been received; however, complaints of a similar nature referencing aspiration/suction issues have been received and the most likely root cause is an error during the supplier¿s manufacturing process of the blue aspiration luer on the system manifold.It has been found that the blue luer on the manifold is not creating a complete seal resulting in air flow.
|
|
Event Description
|
Additional information provided.The decrease of aspiration occurred during irrigation/aspiration phase.Additional information has been requested and received.This is one of five reports being filed for this facility.This report refers to the event that happened on (b)(6) 2016.
|
|
Manufacturer Narrative
|
Evaluation summary: a review of the device history record indicated the order was built to specification.The customer did not retain a sample for this complaint report; therefore, visual inspection or functional testing could not be conducted.The system operator's manual provides the following in regard to loss of aspiration/suction issues: insufficient aspiration probable cause: loose blue luer fittings.Damaged o-ring (ultraflow irrigation/aspiration handpiece only).Clogged tip.Kinked or damaged tubing.Cracked blue fitting.Recommended solutions: reconnect securely inspect o-ring and replace, as necessary.Flush tip with sterile water or bss sterile irrigation solution.Retest.Replace tip.Retest.Check tubing and/or replace cassette.Check fitting and/or replace cassette.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.
|
|
Search Alerts/Recalls
|