BARD PERIPHERAL VASCULAR, INC. IMPRA VASCULAR GRAFT Back to Search Results
Catalog Number F7006TWS
Device Problem Torn Material (3024)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer the lot number for the device has been provided.A review of the device history records is currently being performed.The device has been not returned to the manufacturer for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the health care professional was allegedly removing the beads on the graft gently, when the graft tore.Reportedly, the tore section was cut off and the remainder was used to complete the procedure.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: the date of event was during the week of 12/24/15; however a specific date was unobtainable.The lot number was provided and the lot device history records were reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Visual inspection & functional/performance evaluation: the device was not returned; therefore, no visual inspection or functional testing of the device was performed.Medical records review: medical records were not provided for review.Image/photo review: images/photos were not provided for review.Conclusion: the investigation is inconclusive, as a sample was not returned for evaluation.The root cause could not be determined based upon available information.It is unknown whether procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not remove the external spiral support (beading) from any centerflex¿ graft or endflex¿ graft.Attempts to remove the beading may damage the graft wall.If damage occurs, discard the graft.Precautions: when removing the external spiral support (beading) of impra flex¿ grafts, the beading must be removed slowly and at a 90° angle to the graft.Rapid unwinding and/or removal at less than a 90° angle may result in graft damage.Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall.If damage occurs, that segment of the graft should not be used.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRA VASCULAR GRAFT
Type of Device
VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 west 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 west 3rd st.
tempe AZ 85281
Manufacturer Contact
judith ludwig
1625 west 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key5394316
MDR Text Key37431083
Report Number2020394-2016-00131
Device Sequence Number1
Product Code DYF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K983064
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2016
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue NumberF7006TWS
Device Lot NumberVTXH2522
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient Weight77
-
-