Results: the penumbra system 3max reperfusion catheter (3max) was fractured approximately 43.0 cm from the hub, and kinked approximately 38.0 cm from the distal tip.Conclusions: evaluation of the returned device revealed that the 3max was fractured and kinked.These types of damage typically occur due to improper handling during preparation for use.If the 3max is removed from its packaging hoop with force at an angle, or otherwise manipulated at an angle, these types of damage may occur.These devices are 100% visually evaluated during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|