MAQUET CARDIOPULMONARY AG HLM TUBING SET W/SOFTLINE COATING; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HQV 32902
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).The product was investigated in the laboratory of the manufacturer.The blue filter cap was not easy to open.It was determined that the blue filter, sitting on the spike, was not removable from the spike by opening the cap, turning or pulling on it.Based on this a confirmation of the failure was not possible.This data will be handled through a designated maquet trending process.If a trend occurs, it will be escalated to quality assurance management for review and determination if further investigation is necessary.Due to this no further action will be completed at this time.
 
Event Description
It was reported that the blue filter cap of the prime line was loose.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/SOFTLINE COATING
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
MICHAEL CAMPBELL
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
MDR Report Key5398413
MDR Text Key37728245
Report Number8010762-2016-00049
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K090533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2017
Device Model NumberHQV 32902
Device Catalogue Number70102.6439
Device Lot Number92159142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-