ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during biomed testing the device displayed "pacer fault 115" and "pacer disabled" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The customer was contacted for return of the device.The device has not been returned to zoll for evaluation.
 
Manufacturer Narrative
The device was not returned to zoll medical corporation; the device's pace/defib engine was removed and returned.The malfunction was duplicated and attributed to a faulty inductor on the pace/defib engine.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5407292
MDR Text Key37666260
Report Number1220908-2016-00152
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K060559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-