Model Number R SERIES |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing the device displayed "pacer fault 115" and "pacer disabled" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The customer was contacted for return of the device.The device has not been returned to zoll for evaluation.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation; the device's pace/defib engine was removed and returned.The malfunction was duplicated and attributed to a faulty inductor on the pace/defib engine.Analysis for reports of this type has not identified an increase in trend.
|
|