• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. SIGMA HP UNI INS SZ2 7MM LM/RL; KNEE TIBIAL BEARING/INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. SIGMA HP UNI INS SZ2 7MM LM/RL; KNEE TIBIAL BEARING/INSERT Back to Search Results
Catalog Number 102453207
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6) 2016 (b)(4), (b)(6) 2020.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
The poly insert would not lock into the tray.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The received device was forwarded to commercialized product development for evaluation.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Review of the device history records did not reveal any related manufacturing deviations or anomalies on the provided product and lot combination.The damage is consistent with unsuccessful attempts to mate the tibial insert and tray.The root cause, however, cannot be definitively determined with the information provided.Corrective action was not indicated.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ2 7MM LM/RL
Type of Device
KNEE TIBIAL BEARING/INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5408254
MDR Text Key37728586
Report Number1818910-2016-12622
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK070267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 01/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Catalogue Number102453207
Device Lot Number722692
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-