It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.Also, it was reported that there was a double balloon breach of the catheter.The patient experienced hypotension and a ventricular heart rhythm.It was noted that the patient stabilized after supportive measures and the nitrous oxide dissipated as the patient recovered.The balloon catheter was removed and replaced.The procedure was completed with the cryo console.No further patient complications have been reported as a result of this event.
|
Product event summary: the balloon catheter and data files were returned and analyzed.The data files confirmed the system notices that indicated that the safety system detected fluid in the catheter and stopped the injection (50005) and a compromised outer vacuum was detected simultaneously at application six.Also, six applications were performed using the catheter.Furthermore, the data files showed a high baseline flow at the beginning of the inflation.Visual inspection of the catheter showed that the device was full of blood.The inner and outer balloons were ruptured around the equator.Also, the distal ends of the balloons were folded backwards.There was a kink at the tip that induced while trying to clear the tip and the guide wire lumen material was extremely soft.The proximal bonding outside diameter (od) was within specification.The catheter shaft was kinked at two locations.The first kink was 2.12 inches distal from the strain relief.The second kink was 1.5 inches proximal from the distal shaft.Also, the smart chip verification indicated that the catheter was used for six injections.The guide wire lumen was pressurized and no breach of the guide wire lumen was observed.In conclusion, the reported issue was confirmed through testing and through data analysis.The catheter failed the visual inspection due to double balloon rupture, guide wire lumen pinch and shaft kink.Performance testing and failure analysis were not conducted due to the condition of the catheter.
|