ROCHE DIAGNOSTICS COAGUCHEK XS PLUS KIT USA; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 04800842160
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2016
Event Type  malfunction  
Manufacturer Narrative
Na.
 
Event Description
The customer called stating the coaguchek xs plus meter (serial number (b)(4)) was powering off on its own during a test.During troubleshooting the call agent advised the customer to clean the battery contacts.The customer stated one of the contacts looks burnt.There was no adverse event.The meter was requested to be returned and a replacement meter was sent.
 
Manufacturer Narrative
The meter was evaluated and the investigation results showed that there were no burn marks on the battery contacts.One of the battery contacts was bent but it did not influence the functionality of the meter.The meter did not show any failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PLUS KIT USA
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5416808
MDR Text Key38249839
Report Number1823260-2016-00120
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04800842160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-