Catalog Number 6605526
Device Problems Complete Blockage (1094); Fluid/Blood Leak (1250)
Patient Problem Not Applicable (3189)
Event Date 01/19/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument on (b)(4) 2016.The fse found an obstruction at the probe rinse block.He removed the obstruction and the instrument ran without any leaks.The repairs were verified per established service procedures.(b)(4).
 
Event Description
The customer observed a fluid leak of 3 ml from a coulter hmx hematology analyzer with autoloader.The leak was not contained within the instrument and no error messages were reported by the customer at the time of the occurrence.The customer was wearing personal protective equipment consisting of a laboratory coat, goggles and gloves at the time of the event and there was no report of injury or direct exposure to the leak.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
kendall FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue
32-s08
kendall, FL 33196
3053802031
MDR Report Key5418013
MDR Text Key37735813
Report Number1061932-2016-00079
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K010765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605526
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-