• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SIMPLEX P - CE ABC FD 10-PK; BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SIMPLEX P - CE ABC FD 10-PK; BONE CEMENT Back to Search Results
Catalog Number 61969010
Device Problems Break (1069); Out-Of-Box Failure (2311); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 01/14/2016
Event Type  malfunction  
Manufacturer Narrative
The event involves a device that is not cleared for sale in the u.S., but similar device is commercially available in the u.S.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
The customer reported that "the antiotics was broken inside the packaging.".
 
Manufacturer Narrative
An event regarding packaging damage involving simplex bone cement with tobramycin was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as no devices were returned and no photographs were provided.Medical records received and evaluation: not performed as there was no patient involvement.Device history review: review of the batch manufacturing record indicates that this batch was manufactured and shipped to stock with no reported discrepancies.Complaint history review: review determined that there were no other similar reported events for the lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as device return and/or photographs are needed to complete the investigation for determining a root cause.A capa trend analysis was conducted for the reported failure mode and concluded simplex packaging damage may result from other factors not necessarily related to the product.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
The customer reported that "the antiotics was broken inside the packaging.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLEX P - CE ABC FD 10-PK
Type of Device
BONE CEMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5419048
MDR Text Key38288883
Report Number0002249697-2016-00361
Device Sequence Number1
Product Code LOD
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nursing Assistant
Type of Report Initial,Followup
Report Date 01/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2017
Device Catalogue Number61969010
Device Lot NumberBGW022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-