MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problems Failure to Sense (1559); Moisture or Humidity Problem (2986)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/03/2014
Event Type  malfunction  
Manufacturer Narrative
This medwatch report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.The customer was sent a pressure xdcr and a proximal sense line.As of this date the parts have not been received for evaluation.When they have been received and evaluated a supplement will be filed.
 
Event Description
(b)(4) was received by customer advocacy june 9, 2014.The report stated: two sensormedics high frequency ventilators (models 3100a and 3100b) have failed in use in recent months.The investigation found water ingress into the proximal airway line which caused the pressure sensors to fail.I have comments from the manufacturer that contradict each other and pictures of the internal circuit of the machine if required.Manufacturer's response: technical support #1: "water ingress frequently occurs on the 3100b because it uses a shorter circuit.Dependent on patient y position."technical support #2 (from manual): "a trickle flow of dry gas from the blender flows constantly from the 3100b to the patient y to keep water vapor from even partially obstructing this pressure sensing pathway".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5419729
MDR Text Key38435228
Report Number2021710-2016-03053
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2006
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse