(b)(4) was received by customer advocacy june 9, 2014.The report stated: two sensormedics high frequency ventilators (models 3100a and 3100b) have failed in use in recent months.The investigation found water ingress into the proximal airway line which caused the pressure sensors to fail.I have comments from the manufacturer that contradict each other and pictures of the internal circuit of the machine if required.Manufacturer's response: technical support #1: "water ingress frequently occurs on the 3100b because it uses a shorter circuit.Dependent on patient y position."technical support #2 (from manual): "a trickle flow of dry gas from the blender flows constantly from the 3100b to the patient y to keep water vapor from even partially obstructing this pressure sensing pathway".
|