• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC358XL
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2016
Event Type  malfunction  
Manufacturer Narrative
The device has not been made available for evaluation at this time.Should further information or the device become available for evaluation, a follow-up report will then be issued.
 
Event Description
Dealer called and stated that the consumer alleges the chair started smoking so they called the fire department (911) who took the chair outside.
 
Manufacturer Narrative
The device was returned and evaluated.The evaluation determined the failure mode to be a diode on the pcb.
 
Event Description
Dealer called and stated that the consumer alleges the chair started smoking so they called the fire department (911) who took the chair outside.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
182 susquehanna ave
n/a
exeter PA 18643
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
182 susquehanna ave
n/a
exeter, PA 18643
5706555574
MDR Report Key5428217
MDR Text Key38474770
Report Number2530130-2016-00014
Device Sequence Number1
Product Code INO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070950
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberLC358XL
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-