Catalog Number 5C4471R |
Device Problems
Smoking (1585); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Should additional information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was a power failure with the homechoice (hc) during patient setup.The care giver turned the hc on and heard a popping noise, then saw smoke come from the hc.The technical service representative would swap the hc.There was patient involvement.There was no patient injury or medical intervention indicated at the time of the report.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned and the evaluation is complete.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Internal and external inspection was performed and issues were noted.The power harness connector is overheated at the accomp board side.The homechoice device received a returned instrument testing evaluation (rite).The device received rite functional and electrical testing.The device was received inoperative for no power, and then failed the power up verification step.The device passed the rite electrical test.Popping noise with smoke was verified by visual observation.The direct cause for popping noise with smoke was determined to be poor connection at the accomp board header to power harness interface connector j1 by inspection.The reported condition was verified.The product analysis lab has identified the entire device to be scrapped.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|