• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED Back to Search Results
Model Number M41SR20B
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A returned was issued and the product was evaluated upon receipt.A follow up will be filed if/when any additional information is provided.On (b)(6) return was evaluated and the result was: discovered during bench test that the brake handle is loose allowing the micro switch to be triggered.Brake engages/disengages properly during bench test.Wiggled wires and brake handle with no failure.
 
Event Description
Dealer is stating that the chair was given motor faults.Return was evaluated and the result was: discovered during bench test that the brake handle is loose allowing the micro switch to be triggered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
Type of Device
WHEELCHAIR, POWERED
Manufacturer (Section D)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH  215121
Manufacturer (Section G)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH   215121
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5436454
MDR Text Key38800263
Report Number3008262382-2016-00115
Device Sequence Number1
Product Code ITI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberM41SR20B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-