The device was implanted via v-p shunt to a (b)(6) female.Doi and the initial setting pressure are unknown.After implantation, the patient had a minor headache.The surgeon tried to change the setting pressure, but it could not be changed even by using neodymium.The device was replaced with the same new product at 90mmh2o on (b)(6) 2016, and the patient's condition improved.The indicator of the valve is positioned between 30mmh2o and 200mmh2o, so the surgeon suspects the cam dislodgement or the valve becoming stuck at the home position due to biological debris.
|
Upon completion of the investigation it was noted that the position of the cam when valve was received was 30mmh2o.The valve was visually inspected: no defects were noted.The valve was tested for programming.With programmers 82-3126 sn (b)(4) and 82-3190 sn (b)(4), the valve failed the test, during the programming process the cam mechanism did not move.The valve was irrigated with purified water.An occlusion was noted at the inlet of the ruby ball.Therefore a fine needle was inserted into the flow opening of the valve inlet under the ruby ball and its seat, the ruby ball was manually popped free from its stuck position using light force.The valve was irrigated again, no occlusion was noted.The catheters were irrigated with purified water, no occlusion was noted.The valve was dried.The valve was leak tested, no leaks were found.The valve was reflux tested, the valve passed the test.The valve was retested for programming.With programmer 82-3126 sn (b)(4) and 82-3190 sn (b)(4), the valve failed the test, during the programming process the cam mechanism did not move.The valve was then pressure tested at 30mmh2o, the valve failed the test.The valve was dismantled and was examined under microscope at appropriate magnification: biological debris was found on the spring, on the spring pillar, on the ruby ball, on the seat of the ruby ball, on the cam mechanism, on the cam mechanism pillar, and on the base plate.The cam magnets were also controlled.The magnets passed.The lot history record was reviewed for completeness during the release process to inventory.At that time based on the fact that no discrepancies were noted for the products being accepted, they were released to stock on the 22nd june 2006.The root cause of the problem reported by the customer is due to biological debris found on the spring, on the spring pillar, on the ruby ball, on the seat of the ruby ball, on the cam mechanism, on the cam mechanism pillar, and on the base plate.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|