MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2016
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the sidecom printed circuit board (pcb) being pushed in the board.Per the hill-rom service manual the bed in order to maintain its characteristics and performance.Make sure the head and foot side rails are not bent or twisted.Make sure the latch mechanism operates correctly should be subject to an effective maintenance program.An annual service of the bed is advised.An audible click must be heard.Remove the side rail cover, and make sure the mounting screws are tight.Inspect the cable routing for pinching, binding, and damage.Make sure all functions on the caregiver control operate correctly.Repair or replace the side rail as necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the sidecom pcb to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the nurse call function was working intermittently.The bed was located at the account in 3332.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key5446876
MDR Text Key39001533
Report Number1824206-2016-00091
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-