MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1011707-38
Device Problems Detachment Of Device Component (1104); Failure to Advance (2524); Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/21/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation summary: (b)(4).The device was returned for analysis.The reported physical property issue was able to be confirmed.The reported failure to advance the device was unable to be replicated in a testing environment as it was based on operational circumstances.Based on a visual and dimensional inspection of the returned device, there is no indication of a product quality issue with respect to manufacture, design or labeling.Based on the information provided, the reported difficulties appear to be related to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.Based on the information reviewed, there is no evidence to indicate the presence a potential quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that during the procedure the 2.5 x 38 mm and 2.25 x 28 mm xience prime stent could not cross the lesion and buckled and were damaged.There was a large piece of calcium in the vessel.No additional information was provided.Return device analysis for the 2.5 x 38 mm xience prime revealed the support mandrel was separated at the distal end of the bayonet and was not returned.There was a tear in the outer member distal to the bayonet, for a length of 3mm.No.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5452107
MDR Text Key39028016
Report Number2024168-2016-01042
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 02/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/13/2018
Device Catalogue Number1011707-38
Device Lot Number5100841
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
-
-