• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON 22MM DUAL HTD CIRC KIT; RESPIRATORY GAS HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON 22MM DUAL HTD CIRC KIT; RESPIRATORY GAS HUMIDIFIER Back to Search Results
Catalog Number 870-35KIT
Device Problems Loss of or Failure to Bond (1068); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A conclusion code could not be chosen.The complaint was confirmed, but the root cause is unknown.The device history record of batch number (b)(4) that belongs to catalog number 870-35kit has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Upon receipt the circuit was visually examined for any signs of damage, abuse/misuse.It was noted that the connector on the expiratory side of the limb, which attaches to the filter of the ventilator was loose and easily removed.The complaint has been confirmed.Functional inspection - when the connector was pushed firmly back onto the circuit tubing it stayed in place and would not detach without using a significant amount of pulling force.The investigation revealed a loose connector on the expiratory side of the limb.Other remarks: root cause cannot be determined.There are several speculative scenarios that should be mentioned regarding this complaint.First, the complaint description states that the connectors became unglued.These connectors are not glued.They are press fitted onto the circuit with an alcohol lubricant/sealant.Secondly, this connector is handled more frequently than the other connectors because it is attached to the ventilator filter which is changed daily.This end of the expiratory limb must be detached in order to change the ventilator filter which requires more opportunities for this connector to be handled and work loose.No further action required.
 
Event Description
The customer alleges that the circuit came unglued where the expiratory corrugated tubing connects to the hard plastic piece.The circuit was in use for approximately 22 days.No patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON 22MM DUAL HTD CIRC KIT
Type of Device
RESPIRATORY GAS HUMIDIFIER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
margie burton, rn
3015 carrington mill blvd
morrisville, NC 27560
9194334965
MDR Report Key5453784
MDR Text Key38775219
Report Number3004365956-2016-00130
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number870-35KIT
Device Lot Number74M1500851
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HAMILTON G5 VENTILATOR
-
-