Age at time of event: 18 years or older.Device evaluated by manufacturer: the device was returned for analysis.An empty opened tray was returned for the reported.During visual analysis it was noted that a section of the (b)(6) lid was still attached to the tray, indicating that the package was difficult to open.The (b)(6) lid was torn and frayed, however no fibres/particles were found inside the tray.As the encore device was not returned with the packaging the presence of fibers on the device could not be confirmed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is user preference issue as device meets specification however the user is dissatisfied with the function, performance, and the appearance of the product.(b)(4).
|
It was reported that the sterility was compromised.A 26 encore balloon catheter inflation device was selected for use.During preparation, when the device was removed from the box, it was noted that the bonding part of the plastic case containing the device was unable to peel off appropriately and paper inside the cover caused fibers, making the device unclean.The procedure was completed with another of the same device.No patient complications were reported and patient condition was good.
|