MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 MEC WIDEBAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 MEC WIDEBAND Back to Search Results
Catalog Number 36302
Device Problem Sticking (1597)
Patient Problems Erythema (1840); Pain (1994); Burning Sensation (2146); Discomfort (2330); Blood Loss (2597)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the patient is unable to easily remove the mec because the adhesive is too sticky.Patient has tried to use a warm cloth, as directed, to loosen the adhesive, but is still having difficulty.Patient states that he has used witch hazel and alcohol as alternate measures to try to remove the adhesive, but ultimately has to use his fingernails.As a result, the patient has experienced pain, redness, small amount of bleeding, burning sensation and soreness.
 
Manufacturer Narrative
Received 1 used male external catheter only.Visual inspection noted no obvious defects.Used samples can not be tested for adhesive peel strength.No packaging was returned with the product.The lot number is unknown therefore the device history record could not be reviewed.The complaint was inconclusive due to poor sample condition.The instructions for use state the following: "the self-adhering male external catheter is designed for the management of male urinary incontinence.Do not use on irritated or compromised skin.Do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions on how to apply the catheter: wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Open package at perforation.To remove plastic insert, squeeze catheter at the top of the white cone and pull to release.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions to remove the catheter gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.Non-sterile.Single use device.Use by date.Not made with natural rubber latex." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEC WIDEBAND
Type of Device
MEC WIDEBAND
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5462097
MDR Text Key39446554
Report Number1018233-2016-00203
Device Sequence Number1
Product Code NNX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2020
Device Catalogue Number36302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-