Engineering analysis: the complaint details provided indicate that the stent dislodged off the balloon within the introducer sheath.The returned device was removed from the packaging and inspected to determine the cause of the dislodgement.Upon inspection, the icast stent was still crimped onto the balloon and was not loose on the balloon.The stent was properly positioned between the two radiopaque ro marker bands.The diameter of the stent was measured and was 2.4mm.This diameter is indicative of a properly crimped stent and is in line with the rest of the measurements documented within the quality inspection records.The introducer sheath used in the case was not returned.To determine if the icast delivery system and stent performed properly the returned device was prepped per the instructions listed in the instructions for use(ifu).The balloon was inflated to the nominal inflation pressure of 8atm.The stent opened properly within the ro marker band and the stent did not move.The pressure was then increased to the rated burst pressure of 12atm as specified on the product label.The product performed as expected.There are no indications that the stent was dislodged.The return details provided indicate that the stents intended target was the celiac trunk.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath.Ability to deploy the stent at nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.Balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm) manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs result: all 59 quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.All samples passed through the introducer sheath without any stent dislodgments.During the final lot qualification data shows that all 59 test samples were able to pass through the 7fr introducer sheath without issue.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.
|